Skip to main content

Article Filters

Re-use, Cleaning and sanitization of IsoClear™ Affinity Adsorbents

Published date: 12 December 2023

Back to Article Listing

Isoagglutinins are antibodies (IgM and IgG isotypes) which recognise and bind to type A or B blood group antigens. The presence of isoagglutinin antibodies in plasma-derived products (particularly plasma for transfusion and IVIG) can give rise to serious side effects, such as haemolysis, depending upon the type of isoagglutinin present (A or B) and the blood type of the recipient. Historically, individuals of AB blood type have been used as ‘Universal Plasma’ donors, as their plasma does not contain isoagglutinin antibodies. However, AB is the rarest blood group, possessed by a very low percentage of the population. Thus, AB plasma is in great demand and only available in limited quantities.

Astrea Bioseparations have developed two affinity chromatography adsorbents (IsoClear A™ & IsoClear B™) comprising of immobilized trisaccharide blood group antigens (A & B). These antigens are chemically synthesized and covalently attached to a bio-compatible polymethacrylate support matrix. IsoClear™ adsorbents are highly selective for the removal of isoagglutinin antibodies (IgM and IgG isotypes), enabling cost effective and efficient reduction of isoagglutinin titre for plasma and plasma derived products.

IsoClear A™ & IsoClear B™ are applicable to most blood derived compositions with excellent protein recovery including protective antibodies, coagulation and thrombotic factors. The adsorbents are easy to pack with excellent flow properties and are robust, long-life resins, capable of multiple cycles if the appropriate cleaning and sanitization solutions are used.
In this application note, appropriate cleaning and sanitization methods for IsoClear A™ & IsoClear B™ are investigated and described.

Materials & Methods

Re-use study

A scale-down reuse study was performed on an automated chromatography workstation using a 2 mL IsoClear B™ column challenged with 50 CV (100 mL) of plasma per cycle. Column cycling was continued for 50 cycles. The base-line chromatography conditions used are summarized in the table below:

Table 1: Chromatography conditions for the IsoClear B™ re-use study using plasma.

 

CIP study 1

A wide range of conditions were investigated both in a 96-column PuraPlate™ format as well
as individual column format (data not shown). The final conditions identified a combination of denaturants and reducing agents which provided improved cleaning.

Table 2: Chromatography conditions for the IsoClear™ B CIP study 1 using plasma.

 

Different guanidinium salts (guanidine hydrochloride, guanidine thiocyanate and
guanidine carbonate) in a buffer of 20 mM Tris, 1 mM EDTA, 100 mM DTT pH 8.5 were evaluated alongside the current CIP condition (20% ethanol, 1 M acetic acid). Table 2 summarises the chromatography conditions used. Each column had 3 process cycles carried out prior to resin sampling.

CIP study 2

A study to investigate two different reducing agents (100 mM DTT versus 10 mM TCEP) was
performed with the result that the type of reducing agent did not appear to affect the performance of cleaning–in–place as long as the pH was mildly alkaline (data not shown). For CIP study 2, TCEP was selected as the thiol reducing agent. An investigation into varying levels of guanidine hydrochloride from 1 M to 6 M was carried out. 3.5 g of IsoClear™ was incubated (batch mode) for 1 hour with gentle mixing in 350 mL plasma. Post incubation, the resin was captured in a filter, washed with equilibration buffer and equally distributed into a 96-column PuraPlate™. The conditions for the study are highlighted in Table 3.

Table 3: Chromatography conditions for the IsoClear™ B CIP study 2.

 

The 6 M Urea solution in table 3 was prepared in 100 mM DTT, 1 mM EDTA, 20 mM Tris pH 8.5.

 

Peracetic acid investigation

 A scale-down reuse study was performed an automated chromatography workstation and a 2.3-mL IsoClear B™ column challenged with 15 mL of 0.1 M peracetic acid with a contact time of 20 minutes. The re-use study was continued for 50 cycles. The chromatography conditions are summarized in Table 4.

 Table 4: Chromatography conditions for the IsoClear B™ peracetic acid cycling study.

 

After 50 cycles of peracetic acid sanitisation, a 1-mL sample was removed from the column and analysed using a carbohydrate assay to determine ligand density. The remaining resin was repacked and challenged with 100 mL of 1-μm filtered plasma using the conditions set out in Table 2.

Stability study

An incubation study was performed with different sanitizing solutions: PAB (120 mM phosphoric acid, 167 mM acetic acid, 2.2% benzyl alcohol, 0.2% N,N-Dimethyldodecylamine N-oxide (DDAO) as well as with 3 M guanidine hydrochloride, 100 mM DTT, 1 mM EDTA, 25
mM tris, pH 8.5. A slurry of IsoClear A™ was made and stored continuously for 4 weeks in the stated solutions (roughly equivalent to >150 cycles exposure with a 4-hour contact time per cycle). The resin was washed and analysed using Astrea Bioseparations' standard QC release test methods including a ligand density assay and functionality assay (a static binding test identifying the
depletion of IgM isoagglutinins from a known level of a spiked solution (titre range: 1/256 to 1/512)).

Optimized conditions

A scale-down reuse study was performed on an automated chromatography workstation using a 2-mL IsoClear B™ column challenged with 50 CV (100 mL) of plasma per cycle. Column cycling was continued for over 15 cycles, using the optimized conditions highlighted from the CIP study. The chromatography conditions used are summarized in Table 5.

Table 5: Chromatography conditions for the IsoClear A™ >15 cycle re-use study using plasma with the optimized CIP process.

 

Process samples were analysed by Nephelometry and agglutination assay, and resin removed from the column to identify residual fouling on the column.

Results & Discussion

Re-use Study

50 use cycles were successfully completed for IsoClear™ affinity resin. The chromatograms (Figure 1) show typical profiles for a negative binding step with the majority of the plasma proteins passing through the column and small elution and CIP peaks (containing isoagglutinins). The Load and Non-bound fractions were analysed using a modified Coombs test (Figure 2) demonstrating the clearance of isoagglutinins from 1/64 in the load to 1/2 in the non-bound fraction. This drop in titre is equivalent to ~ 97% removal of isoagglutinins.

Figure 1: Chromatograms (UV overlays) of every 10 runs of the reuse study. 100 mL human plasma loaded on to a 2 mL column. (NB – Non-bound; E – Elution; CIP – Clean-in-place).

 

The recovery of IgG was monitored by nephelometry and no discernible losses were measured over the course of the experiment (~ 100% recovery of IgG after run 51).

Figure 2: Modified Coombs agglutination assay of the Load and Non-bound samples from plasma using IsoClear B™. The load demonstrated a titre of 1/64 and the non-bound, a titre of 1/2. All samples were tested in duplicate.

 

CIP Study 1

The majority of the CIP solutions tried did not have any significant effect on the cleanliness of the resin (data not shown). The investigation into different denaturants, in particular guanidine salts, showed a significant improvement in the removal of fouling from the resin compared to the baseline method (1 M acetic acid, 20% ethanol). 3 M solutions of guanidine hydrochloride, thiocyanate and carbonate in the presence of a reducing agent were used to clean plasma-fouled IsoClear™ resin (100 mM DTT, 1 mM EDTA, 20 mM Tris, pH 8.5) for 1 hour, followed by re-equilibration in starting buffer. Resin samples were removed from the top of the column for SDS-PAGE analysis (Figure 3).

Figure 3: SDS-PAGE of the resin samples from the CIP study 1 (post three process cycles). Lane: 1. SeeBlue +2 molecular weight marker; Lane 2: 20% ethanol, 1 M Acetic acid; Lane 3: 3 M guanidine hydrochloride; Lane 4: 3 M guanidine thiocyanate; Lane 5: 3 M guanidine carbonate. All samples were prepared with reducing agent and loaded neat (20 μL/well).

 

CIP Study 2

A study was performed to investigate 100 mM DTT versus 10 mM TCEP, both in the presence of a denaturant. There was no significant observable difference between the two reducing agents (data not shown). From the first CIP study, guanidine HCl was selected as the most appropriate CIP solution. A range of solutions from 1 M to 6 M guanidine HCl (in the presence of  TCEP in a Tris buffer) were tested against fouled IsoClear™ in a PuraPlate format (along with 6 M urea and 20% ethanol, 1 M acetic acid as a control).

IsoClear™ samples from each CIP condition were subjected to reducing SDS-PAGE and analysed by densitometry to determine the extent of fouling removal.

The results shown in Figure 4 compare the level of residual adsorbent fouling, as well as the
percentage clearance of fouling in comparison to adsorbent which had not been subjected to a clean-in-place procedure.

Figure 4: Graph depicting the residual fouling of the resins after incubation of various CIP solutions. 1 M through to 6 M guanidine HCl (in 10 mM TCEP, 25 mM tris, pH 8.5 buffer), 6 M Urea and the control solution 1M acetic acid, 20% ethanol. The percentage clearance of foulants from the pre-CIP column is highlighted on the graph.

 

As the concentration of guanidine HCl increases, there is an observed decrease in the level of fouling. 1 M guanidine appears to remove only about 58% of the foulants from the column, whereas at 3 M guanidine hydrochloride and beyond the clearance of fouling is ≥ 99%. The use of 3 M guanidine hydrochloride in combination with 10 mM TCEP represents an improvement in the removal of fouling compared to the base-line 20% ethanol, 1 M acetic acid cleaning solution.

Peracetic acid sanitization

50 cycles were successfully performed on IsoClear B™ using 0.1 M peracetic acid (PAA). Post 50 cycles, the resin was tested for ligand density and functionality. By carbohydrate assay, the ligand density was unchanged at 0.76 mg/mL (the QC release specification is > 0.6 mg/mL). To investigate functionality, a 1.3 mL column was packed, equilibrated and loaded with 80 CV of plasma with an isoagglutinin titre of 1/16. The non-bound fraction was tested using a modified Coombs test and no observable agglutination was detected, showing a successful depletion of isoagglutinins.

Stability study

The stability of IsoClear A™ was investigated by incubating the resin for 4 weeks (> 670 hours) in various CIP and sanitization solutions. The incubated resin samples were rinsed and tested
using the specified QC release tests including bead integrity, ligand density testing and functionality testing.

Exposure to all solutions evaluated (PAB, DDAO and 3 M guanidine hydrochloride, 100 mM DTT, 1 mM EDTA, 25 mM tris, pH 8.5) did not impact product performance (Table 6).

Table 6: Summary results after 4 weeks continuous incubation of IsoClear A™ with CIP and sanitization solutions. A starting time (t) = 0 time point was tested as a reference point. The performance test is based on a IgM spike removal from solution with a titre between 1/256 and 1/512.

 

PAB showed no drop in performance after 4 weeks. There was an apparent drop in ligand density (0.79 mg/ml vs 0.62 mg/ml) but this is still within the release specification for this product. The baseline CIP condition of 20% ethanol, 1 M acetic acid has been previously tested in a stability study and demonstrated no loss of performance (data not shown).

Optimized conditions

 An IsoClear™ resin re-use study was carried out over 15 cycles using the optimized CIP procedure, in comparison to the baseline CIP conditions of 20% ethanol, 1 M acetic acid. The load and non-bound samples were analysed using a modified Coombs test method and a 6 titre clearance was observed from 1/128 to 1/2 or 98% clearance of isoagglutinins (data not shown).

Figure 5 demonstrates the chromatogram overlays for runs 1 and 15 for the optimized CIP conditions, which shows close comparability between the runs.

Figure 5: Chromatograms (UV overlays) of the optimized 15 cycle reuse study. 100 mL human plasma loaded on to a 2 mL column. (NB – Non-bound; E – Elution; CIP – Clean-in-place). Blue trace; run 1; Orange trace, run 15.

 

The optimized CIP condition, demonstrated improved cleaning efficacy in the removal of residual fouling on the column, compared to the baseline CIP (as shown in Figure 3), which allows for multiple cycles without affecting the performance of the resin.

Conclusions

IsoClear™ affinity adsorbents have been specifically designed for the capture and removal of isoagglutinins from plasma and plasma derived feedstocks allowing for a 4 to 6 titre clearance (94% to 98% removal) from plasma. IsoClear™ is a robust, long-life resin, which can be used for multiple cycles using the appropriate CIP and sanitization solutions. 20% ethanol, 1 M acetic acid has been shown to allow IsoClear™ to be re-used for over 50 cycles, however, the extent of cleaning this solution provides is dependent on the feedstock.

Other CIP solutions such as denaturants in combination with reducing agents, in particular 3 M guanidine hydrochloride, 20 mM Tris, 10 mM TCEP, pH 8.5 has been found to be highly effective for cleaning IsoClear™ resins.

Sanitization, critical for bioprocess media, can be performed on IsoClear™ resins using a variety of sanitizing solutions including 0.2% N,NDimethyldodecylamine N-oxide (DDAO), 0.1 M peracetic acid (PAA) and 120 mM phosphoric acid, 167 mM acetic acid, 2.2% benzyl alcohol (PAB), which is an accepted sanitization solution for bioprocess media such as Protein A resins.

IsoClear™ resins are compatible with all of these solutions and are fully functional even after 4 weeks continuous incubation. This demonstrates the robustness of the IsoClear™ products and enables IsoClear™ resins to be used for multiple process cycles (> 50).

IsoClear™ resins have high binding capacities for isoagglutinins and can be used to remove A and B isoagglutinins from complex feedstocks (plasma and plasma derivatives). These resins can be used individually or as a combination of IsoClear A™and IsoClear B™, can be operated at plasma:resin ratios of up to 200 to 1 with very short residence times (≤ 3 minutes) and high flow rates (400 cm/h at 1 bar in a 10 cm diameter with a 12 cm bed height). As a result, IsoClear™ affinity adsorbents are perfectly suited to the production of universal plasma and for reducing the isoagglutinin titre of IVIG preparations.

Your basket

Your basket is empty. Continue shopping to add products to your basket.

Search our catalogue

Call Centre

Request Quote

Please enter the quantity of products you want a quote for

Please enter the date you need these products byPlease enter a date in the format DD/MM/YYYY

View Content

Please complete the form below to receive the requested content from Astrea Bioseparations. We will use the data you provide to send you relevant updates, special offers, and product-related information. See our Privacy Notice for details on how we use personal data.

If you have an account please log in to gain access to content. You can apply for an account here

Please enter a First Name

Please enter a valid First Name, the maximum length is 50 characters.

Please enter a Last Name

Please enter a valid Last Name, the maximum length is 50 characters.

Please enter a Company Name

Please enter a valid Company, the maximum length is 100 characters

Please enter a valid Email AddressPlease enter a valid Email AddressThe Email Address entered is already registered, please sign in with the Email Address or enter a different one

Please select a valid Telephone Number, the maximum length is 30 characters

Please enter a valid Telephone Number consisting only of the following characters and spaces ( ) + 0 1 2 3 4 5 6 7 8 9

Please select a State / Province / Territory

Please select a valid State / Province / Territory, the maximum length is 150 characters

Please select a Country

Data entered into form will be stored in your session (not in a cookie) to avoid having to re-enter in future form submissions.

Re-use, Cleaning and sanitization of IsoClear™ Affinity Adsorbents

Published date: 12 December 2023

Back to Article Listing

Isoagglutinins are antibodies (IgM and IgG isotypes) which recognise and bind to type A or B blood group antigens. The presence of isoagglutinin antibodies in plasma-derived products (particularly plasma for transfusion and IVIG) can give rise to serious side effects, such as haemolysis, depending upon the type of isoagglutinin present (A or B) and the blood type of the recipient. Historically, individuals of AB blood type have been used as ‘Universal Plasma’ donors, as their plasma does not contain isoagglutinin antibodies. However, AB is the rarest blood group, possessed by a very low percentage of the population. Thus, AB plasma is in great demand and only available in limited quantities.

Astrea Bioseparations have developed two affinity chromatography adsorbents (IsoClear A™ & IsoClear B™) comprising of immobilized trisaccharide blood group antigens (A & B). These antigens are chemically synthesized and covalently attached to a bio-compatible polymethacrylate support matrix. IsoClear™ adsorbents are highly selective for the removal of isoagglutinin antibodies (IgM and IgG isotypes), enabling cost effective and efficient reduction of isoagglutinin titre for plasma and plasma derived products.

IsoClear A™ & IsoClear B™ are applicable to most blood derived compositions with excellent protein recovery including protective antibodies, coagulation and thrombotic factors. The adsorbents are easy to pack with excellent flow properties and are robust, long-life resins, capable of multiple cycles if the appropriate cleaning and sanitization solutions are used.
In this application note, appropriate cleaning and sanitization methods for IsoClear A™ & IsoClear B™ are investigated and described.

Materials & Methods

Re-use study

A scale-down reuse study was performed on an automated chromatography workstation using a 2 mL IsoClear B™ column challenged with 50 CV (100 mL) of plasma per cycle. Column cycling was continued for 50 cycles. The base-line chromatography conditions used are summarized in the table below:

Table 1: Chromatography conditions for the IsoClear B™ re-use study using plasma.

 

CIP study 1

A wide range of conditions were investigated both in a 96-column PuraPlate™ format as well
as individual column format (data not shown). The final conditions identified a combination of denaturants and reducing agents which provided improved cleaning.

Table 2: Chromatography conditions for the IsoClear™ B CIP study 1 using plasma.

 

Different guanidinium salts (guanidine hydrochloride, guanidine thiocyanate and
guanidine carbonate) in a buffer of 20 mM Tris, 1 mM EDTA, 100 mM DTT pH 8.5 were evaluated alongside the current CIP condition (20% ethanol, 1 M acetic acid). Table 2 summarises the chromatography conditions used. Each column had 3 process cycles carried out prior to resin sampling.

CIP study 2

A study to investigate two different reducing agents (100 mM DTT versus 10 mM TCEP) was
performed with the result that the type of reducing agent did not appear to affect the performance of cleaning–in–place as long as the pH was mildly alkaline (data not shown). For CIP study 2, TCEP was selected as the thiol reducing agent. An investigation into varying levels of guanidine hydrochloride from 1 M to 6 M was carried out. 3.5 g of IsoClear™ was incubated (batch mode) for 1 hour with gentle mixing in 350 mL plasma. Post incubation, the resin was captured in a filter, washed with equilibration buffer and equally distributed into a 96-column PuraPlate™. The conditions for the study are highlighted in Table 3.

Table 3: Chromatography conditions for the IsoClear™ B CIP study 2.

 

The 6 M Urea solution in table 3 was prepared in 100 mM DTT, 1 mM EDTA, 20 mM Tris pH 8.5.

 

Peracetic acid investigation

 A scale-down reuse study was performed an automated chromatography workstation and a 2.3-mL IsoClear B™ column challenged with 15 mL of 0.1 M peracetic acid with a contact time of 20 minutes. The re-use study was continued for 50 cycles. The chromatography conditions are summarized in Table 4.

 Table 4: Chromatography conditions for the IsoClear B™ peracetic acid cycling study.

 

After 50 cycles of peracetic acid sanitisation, a 1-mL sample was removed from the column and analysed using a carbohydrate assay to determine ligand density. The remaining resin was repacked and challenged with 100 mL of 1-μm filtered plasma using the conditions set out in Table 2.

Stability study

An incubation study was performed with different sanitizing solutions: PAB (120 mM phosphoric acid, 167 mM acetic acid, 2.2% benzyl alcohol, 0.2% N,N-Dimethyldodecylamine N-oxide (DDAO) as well as with 3 M guanidine hydrochloride, 100 mM DTT, 1 mM EDTA, 25
mM tris, pH 8.5. A slurry of IsoClear A™ was made and stored continuously for 4 weeks in the stated solutions (roughly equivalent to >150 cycles exposure with a 4-hour contact time per cycle). The resin was washed and analysed using Astrea Bioseparations' standard QC release test methods including a ligand density assay and functionality assay (a static binding test identifying the
depletion of IgM isoagglutinins from a known level of a spiked solution (titre range: 1/256 to 1/512)).

Optimized conditions

A scale-down reuse study was performed on an automated chromatography workstation using a 2-mL IsoClear B™ column challenged with 50 CV (100 mL) of plasma per cycle. Column cycling was continued for over 15 cycles, using the optimized conditions highlighted from the CIP study. The chromatography conditions used are summarized in Table 5.

Table 5: Chromatography conditions for the IsoClear A™ >15 cycle re-use study using plasma with the optimized CIP process.

 

Process samples were analysed by Nephelometry and agglutination assay, and resin removed from the column to identify residual fouling on the column.

Results & Discussion

Re-use Study

50 use cycles were successfully completed for IsoClear™ affinity resin. The chromatograms (Figure 1) show typical profiles for a negative binding step with the majority of the plasma proteins passing through the column and small elution and CIP peaks (containing isoagglutinins). The Load and Non-bound fractions were analysed using a modified Coombs test (Figure 2) demonstrating the clearance of isoagglutinins from 1/64 in the load to 1/2 in the non-bound fraction. This drop in titre is equivalent to ~ 97% removal of isoagglutinins.

Figure 1: Chromatograms (UV overlays) of every 10 runs of the reuse study. 100 mL human plasma loaded on to a 2 mL column. (NB – Non-bound; E – Elution; CIP – Clean-in-place).

 

The recovery of IgG was monitored by nephelometry and no discernible losses were measured over the course of the experiment (~ 100% recovery of IgG after run 51).

Figure 2: Modified Coombs agglutination assay of the Load and Non-bound samples from plasma using IsoClear B™. The load demonstrated a titre of 1/64 and the non-bound, a titre of 1/2. All samples were tested in duplicate.

 

CIP Study 1

The majority of the CIP solutions tried did not have any significant effect on the cleanliness of the resin (data not shown). The investigation into different denaturants, in particular guanidine salts, showed a significant improvement in the removal of fouling from the resin compared to the baseline method (1 M acetic acid, 20% ethanol). 3 M solutions of guanidine hydrochloride, thiocyanate and carbonate in the presence of a reducing agent were used to clean plasma-fouled IsoClear™ resin (100 mM DTT, 1 mM EDTA, 20 mM Tris, pH 8.5) for 1 hour, followed by re-equilibration in starting buffer. Resin samples were removed from the top of the column for SDS-PAGE analysis (Figure 3).

Figure 3: SDS-PAGE of the resin samples from the CIP study 1 (post three process cycles). Lane: 1. SeeBlue +2 molecular weight marker; Lane 2: 20% ethanol, 1 M Acetic acid; Lane 3: 3 M guanidine hydrochloride; Lane 4: 3 M guanidine thiocyanate; Lane 5: 3 M guanidine carbonate. All samples were prepared with reducing agent and loaded neat (20 μL/well).

 

CIP Study 2

A study was performed to investigate 100 mM DTT versus 10 mM TCEP, both in the presence of a denaturant. There was no significant observable difference between the two reducing agents (data not shown). From the first CIP study, guanidine HCl was selected as the most appropriate CIP solution. A range of solutions from 1 M to 6 M guanidine HCl (in the presence of  TCEP in a Tris buffer) were tested against fouled IsoClear™ in a PuraPlate format (along with 6 M urea and 20% ethanol, 1 M acetic acid as a control).

IsoClear™ samples from each CIP condition were subjected to reducing SDS-PAGE and analysed by densitometry to determine the extent of fouling removal.

The results shown in Figure 4 compare the level of residual adsorbent fouling, as well as the
percentage clearance of fouling in comparison to adsorbent which had not been subjected to a clean-in-place procedure.

Figure 4: Graph depicting the residual fouling of the resins after incubation of various CIP solutions. 1 M through to 6 M guanidine HCl (in 10 mM TCEP, 25 mM tris, pH 8.5 buffer), 6 M Urea and the control solution 1M acetic acid, 20% ethanol. The percentage clearance of foulants from the pre-CIP column is highlighted on the graph.

 

As the concentration of guanidine HCl increases, there is an observed decrease in the level of fouling. 1 M guanidine appears to remove only about 58% of the foulants from the column, whereas at 3 M guanidine hydrochloride and beyond the clearance of fouling is ≥ 99%. The use of 3 M guanidine hydrochloride in combination with 10 mM TCEP represents an improvement in the removal of fouling compared to the base-line 20% ethanol, 1 M acetic acid cleaning solution.

Peracetic acid sanitization

50 cycles were successfully performed on IsoClear B™ using 0.1 M peracetic acid (PAA). Post 50 cycles, the resin was tested for ligand density and functionality. By carbohydrate assay, the ligand density was unchanged at 0.76 mg/mL (the QC release specification is > 0.6 mg/mL). To investigate functionality, a 1.3 mL column was packed, equilibrated and loaded with 80 CV of plasma with an isoagglutinin titre of 1/16. The non-bound fraction was tested using a modified Coombs test and no observable agglutination was detected, showing a successful depletion of isoagglutinins.

Stability study

The stability of IsoClear A™ was investigated by incubating the resin for 4 weeks (> 670 hours) in various CIP and sanitization solutions. The incubated resin samples were rinsed and tested
using the specified QC release tests including bead integrity, ligand density testing and functionality testing.

Exposure to all solutions evaluated (PAB, DDAO and 3 M guanidine hydrochloride, 100 mM DTT, 1 mM EDTA, 25 mM tris, pH 8.5) did not impact product performance (Table 6).

Table 6: Summary results after 4 weeks continuous incubation of IsoClear A™ with CIP and sanitization solutions. A starting time (t) = 0 time point was tested as a reference point. The performance test is based on a IgM spike removal from solution with a titre between 1/256 and 1/512.

 

PAB showed no drop in performance after 4 weeks. There was an apparent drop in ligand density (0.79 mg/ml vs 0.62 mg/ml) but this is still within the release specification for this product. The baseline CIP condition of 20% ethanol, 1 M acetic acid has been previously tested in a stability study and demonstrated no loss of performance (data not shown).

Optimized conditions

 An IsoClear™ resin re-use study was carried out over 15 cycles using the optimized CIP procedure, in comparison to the baseline CIP conditions of 20% ethanol, 1 M acetic acid. The load and non-bound samples were analysed using a modified Coombs test method and a 6 titre clearance was observed from 1/128 to 1/2 or 98% clearance of isoagglutinins (data not shown).

Figure 5 demonstrates the chromatogram overlays for runs 1 and 15 for the optimized CIP conditions, which shows close comparability between the runs.

Figure 5: Chromatograms (UV overlays) of the optimized 15 cycle reuse study. 100 mL human plasma loaded on to a 2 mL column. (NB – Non-bound; E – Elution; CIP – Clean-in-place). Blue trace; run 1; Orange trace, run 15.

 

The optimized CIP condition, demonstrated improved cleaning efficacy in the removal of residual fouling on the column, compared to the baseline CIP (as shown in Figure 3), which allows for multiple cycles without affecting the performance of the resin.

Conclusions

IsoClear™ affinity adsorbents have been specifically designed for the capture and removal of isoagglutinins from plasma and plasma derived feedstocks allowing for a 4 to 6 titre clearance (94% to 98% removal) from plasma. IsoClear™ is a robust, long-life resin, which can be used for multiple cycles using the appropriate CIP and sanitization solutions. 20% ethanol, 1 M acetic acid has been shown to allow IsoClear™ to be re-used for over 50 cycles, however, the extent of cleaning this solution provides is dependent on the feedstock.

Other CIP solutions such as denaturants in combination with reducing agents, in particular 3 M guanidine hydrochloride, 20 mM Tris, 10 mM TCEP, pH 8.5 has been found to be highly effective for cleaning IsoClear™ resins.

Sanitization, critical for bioprocess media, can be performed on IsoClear™ resins using a variety of sanitizing solutions including 0.2% N,NDimethyldodecylamine N-oxide (DDAO), 0.1 M peracetic acid (PAA) and 120 mM phosphoric acid, 167 mM acetic acid, 2.2% benzyl alcohol (PAB), which is an accepted sanitization solution for bioprocess media such as Protein A resins.

IsoClear™ resins are compatible with all of these solutions and are fully functional even after 4 weeks continuous incubation. This demonstrates the robustness of the IsoClear™ products and enables IsoClear™ resins to be used for multiple process cycles (> 50).

IsoClear™ resins have high binding capacities for isoagglutinins and can be used to remove A and B isoagglutinins from complex feedstocks (plasma and plasma derivatives). These resins can be used individually or as a combination of IsoClear A™and IsoClear B™, can be operated at plasma:resin ratios of up to 200 to 1 with very short residence times (≤ 3 minutes) and high flow rates (400 cm/h at 1 bar in a 10 cm diameter with a 12 cm bed height). As a result, IsoClear™ affinity adsorbents are perfectly suited to the production of universal plasma and for reducing the isoagglutinin titre of IVIG preparations.

Call Centre Product Compare